English: chematic drawing of the FM-AFM setup. A very small tip is mounted on an
oscillating cantilever 1 that is in vicinity of a sample. The oscillation of the cantilever
changes upon interaction between the tip and the sample and is detected with a laser beam
2 focussed on the back of the cantilever. The reflected beam travels via mirrors to a position
sensitive detector (PSD) 3 . The signal of the PSD is amplified by a preamplifier. An
amplitude control 4 measures the amplitude A of this signal and a feedback loop compares
it with a setpoint and determines the amplification (dissipation Γ) of the excitation signal
6 for the cantilever which is fed to a shaking piezo. To measure the current resonance
frequency, a phase-locked loop (PLL) 5 is used. Its voltage-controlled oscillator (VCO)
produces the excitation signal 6 for the cantilever. The detected frequency shift ∆f is
passed to another feedback loop 7 that keeps the frequency shift constant by changing the
distance between the tip and the surface (z position) by varying the voltage applied to the
verbreitet werden – vervielfältigt, verbreitet und öffentlich zugänglich gemacht werden
neu zusammengestellt werden – abgewandelt und bearbeitet werden
Zu den folgenden Bedingungen:
Namensnennung – Du musst angemessene Urheber- und Rechteangaben machen, einen Link zur Lizenz beifügen und angeben, ob Änderungen vorgenommen wurden. Diese Angaben dürfen in jeder angemessenen Art und Weise gemacht werden, allerdings nicht so, dass der Eindruck entsteht, der Lizenzgeber unterstütze gerade dich oder deine Nutzung besonders.
Diese Datei enthält weitere Informationen (beispielsweise Exif-Metadaten), die in der Regel von der Digitalkamera oder dem verwendeten Scanner stammen. Durch nachträgliche Bearbeitung der Originaldatei können einige Details verändert worden sein.