Flow meter error
In flow measurement, flow meter error is typically reported by a percentage indicating non linearity of the device. This can be expressed as either a +/- percentage based on either the full range capacity of the device or as a percentage of the actual indicated flow.

Accuracy of full scale
If a meter’s accuracy is based on a percentage of its full reading capability, the error is a fixed value. For example; an error of 0.5% of full scale, in a 100-gallon per minute (gpm) device is +/- 0.5 gpm. This is the uncertainty all of the time, so as you move away from the full scale capability, the 0.5 gpm error becomes a much larger percentage. At 50 gpm, you are risking a 1% error. At 10 gpm you have a potential 5% error. the flow meters can be affected by the air pressure from the surrounding so it needs to be carefully be observed
Accuracy of reading
If the device has an error expressed as a percentage of the actual flow, then a 0.5% error of 10 gpm is only +/-0.05 gpm; a 10 times better result.[1]