Table of modes of mechanical ventilation

From Wikipedia, the free encyclopedia
This is an old revision of this page, as edited by Josef X Brunner (talk | contribs) at 13:45, 1 April 2025 (I have added an introduction to explain the abbreviations. This is a first step to enhance this table which I find very valuable, especially for clinicians who are bewildered by the jungle of ventilation modes. I do encourage discussion on this entry by a fellow expert.). The present address (URL) is a permanent link to this revision, which may differ significantly from the current revision.

In medicine, mechanical ventilation is a method to mechanically assist or replace spontaneous breathing. Spontaneous breathing requires the rhythmic alteration of inhalation and exhalation to achieve its purpose: exchange of carbon dioxide with oxygen. In disease, the natural process of that rhythmic alteration may be replaced by an external device, a Ventilator, and the clinician has to control the following on the Ventilator to achieve gas exchange:

  1. Composition of gas mixture delivered (Setting: fraction of inspired oxygen, FiO2)
  2. Pressure at the end of exhalation (Setting: Positive End-Expiratory Pressure PEEP)
  3. Control of Breathing Activity
    1. Trigger: Start of inhalation (Settings: time or patient Trigger)
    2. Limit: Time and delivery of gas mixture (Settings: pressure or flow)
    3. Icycle: Cycling from inhalation to exhalation (Settings: time or pressure or volume or flow)
    4. Pcycle: Time and control of exhalation (Settings: time or pressure or flow or resistance)

FiO2 and PEEP settings are common to most Ventilators. Control of Breathing Activity methods, however, can be done in different ways and called "Mode of Mechanical Ventilation"[1]. CAVEAT: Although manufacturers may offer identical Control of Breathing Activity methods, the names of the methods may be different.

A breath by breath trigger, limit, cycling (TLC) classification of the common modes of ventilation. (V = ventilator; P = patient)[2]

Mode Trigger Limit Icycle Pcycle Servo
CMV (VCa, AC) V or P Flow Volume Time no
PCV or PCIRV V or P Pressure Time Time no
SIMV (volume cycled) V or P Flow Volume - no
SIMV (pressure limited) V or P Flow time - no
SIMV+PS (volume cycled) V or P Flow Volume - no
SIMV+PS (pressure limited) V or P Pressure Time - no
CPAP P Pressure Flow - no
CPAP+PS P Pressure Flow - no
BPAP 1 (CPAP+PS) P Pressure Flow - no
BPAP 2 (PCV) V or P Pressure Timed - no
BPAP 3 (CPAP) V Pressure Time - no
BPAP 4 (SIMV) V or P Pressure Time (IV) Flow (IP) - no
BPAP 5 (APRV) V or P Pressure Flow (IP) Time (IV) - no
APRV V or P Pressure Flow (IP) Time (IV) - no
IMPRV V or P Pressure Time - no
PRVC[2] V or P Pressure Time - Yes (VT)
Automode V or P Pressure Flow (IP) Time (IV) - Yes (VT)
Volume Support[2] P Pressure Flow - Yes (VT)
Minimum minute ventilation[2] P Pressure Flow - Yes (VE)
Mandatory rate ventilation[2] P Pressure Flow (IP) Time (IV) - Yes (Vf)
Proportional assist P Pressure/time Flow - Yes (Flow+VE)
Mandatory minute ventilation[2] V or P Flow Time - Yes (VE)