A pulse duplicator is a device used to duplicate the pulsing flow of the human heart and the associated hemodynamics. It is used to research the conditions of heart disease.[1][2] Pulse duplicators can be used to conduct in vitro or ex vivo testing. Common uses include testing new artificial heart valves[3] and simulating procedures like transcatheter aortic valve replacement.

Design

A pulse duplicator replicates parts of the circulatory system. Many pulse duplicators model only half the heart, commonly the left atrium and ventricle, in order to test the mitral and aortic valves.

A common design to model the heart uses a piston to simulate the contraction and expansion of the heart. An alternate design uses a flexible plastic heart model, or an ex vivo heart, and applies hydraulic pressure to induce contraction and expansion.[1]

The circulatory system blood vessels are typically modeled using tubing. Compliance chambers and narrow tubing can be used to model the Windkessel effect.[4]

A pulse duplicator is categorized as a lump parameter model, also known as a Windkessel model, if it uses a limited set of compliance chambers and resistance tubing to model the sum of circulatory compliance and resistance. A pulse duplicator is a wave propagation model if it physically replicates the circulatory system in a more anatomically correct manner. [2]


References

  1. ^ a b Duran, Gunning, McMillan (1964). "A Simple Versatile Pulse Duplicator". Thorax. 19 (6): 503–6. doi:10.1136/thx.19.6.503. PMC 1018869. PMID 14238387.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  2. ^ a b Rodriguez, Reynaldo (December 1, 2017). Redesign and performance evaluation of a cardiac pulse duplicator (Master of Engineering (Mechatronic) thesis). Stellenbosch University. Retrieved June 4, 2024.
  3. ^ Vernon MJ, Mela P, Dilley RJ, Jansen S, Doyle BJ, Ihdayhid AR; et al. (2024). "3D printing of heart valves". Trends Biotechnol. 42 (5): 612–630. doi:10.1016/j.tibtech.2023.11.001. PMID 38238246.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  4. ^ Westerhof N, Lankhaar JW, Westerhof BE (2009). "The arterial Windkessel". Med Biol Eng Comput. 47 (2): 131–41. doi:10.1007/s11517-008-0359-2. PMID 18543011.{{cite journal}}: CS1 maint: multiple names: authors list (link)